Contact Us

Got Questions? We're Here to Help! Contact Us.


California
United States

+1 (510) 899-1340

Welcome to Service AI, your trusted source for accessories, spare parts, and certified pre-owned laboratory instrumentation tailored for scientists and startups. We provide reliable, current equipment at competitive prices, along with essential support for instruments no longer made or backed by their original manufacturers.

Our expertise spans leading brands like Beckman Coulter and Agilent, and we are committed to high-quality standards through certified maintenance, repair, and professional installation services. We believe our installation services provide the confidence and assurance you need when purchasing instrumentation from a reliable and trustworthy source.

At Service AI, we don't conduct the science—we empower scientists with reliable laboratory instrumentation and spare parts that are guaranteed to perform. We trust in scientists' expertise to determine the instrumentation they need to advance their research.

With more than 25 years of experience, our dedicated team prides itself on delivering only the highest quality products. Every instrument we offer has been thoroughly tested and certified by manufacturer-trained engineers. We ensure the functionality of our products with a Full Performance Guarantee and provide proof of performance documentation when applicable.

At Service AI, we also recognize the importance of meticulous shipping and safe delivery. Our products are handled with the utmost care to ensure they arrive in perfect condition. For those who require it, we offer professional installation services to help integrate our equipment seamlessly into your operations.

Purchase our reliable scientific laboratory equipment, and we’ll provide seamless installation service worldwide for your startup or research needs.

Advance your scientific endeavors with confidence: rely on us for recertified, trustworthy instruments at reasonable prices. It’s a smart choice made simple!

Guide to Centrifuge Questions

Service AI support guide for centrifuge questions. Please feel free to email us at support@serviceai.us

Troubleshooting Vacuum Issues in Beckman Coulter Ultracentrifuges: A Comprehensive Guide

Service AI

In an ultracentrifuge, it is crucial to remove air molecules from the 'leak-tight' chamber. To achieve this, Beckman uses a mechanical vacuum pump and diffusion pump in combination. However, in high-performance and high-speed model centrifuges, only a vacuum pump is necessary.

Mechanical vacuum pumps, also known as roughing pumps or forepumps, are the only types that can operate against atmospheric backpressure. The most commonly used mechanical pump is the vane type.

The actual removal of air molecules occurs during four stages of the pumping cycle: intake, transfer, compression, and exhaust.

Vacuum system in an ultracentrifuge

Beckman Coulter ultracentrifuges use vacuum pumps that are characterized by the following features:

 • Two-stage, oil-sealed, rotary vane type

• Maximum vacuum of 20 microns at the inlet

• Pump speed of around 7 cubic feet per minute (cfm)

• Direct drive design with motor in-line with the pump

• Anti-suckback system included

• Mechanical oil stirrer system

• Gas ballast to allow pumping of condensable vapors

• Inlet, outlet, and muffler connections

• Sight glass to indicate oil level

• Easy disassembly into major parts for cleaning and repair

• Motors that can be wired for different voltage requirements

Beckman Coulter ultracentrifuge systems utilize a combination of a diffusion pump and a vacuum pump in series to achieve high levels of vacuum. However, the diffusion pump alone is not capable of exhausting to the atmosphere. As a result, it is exclusively employed in the ultracentrifuge models.

As shown in the diagram below, the Diffusion Pump Assembly consists of the following parts:

• Pump body: a steel cylinder with an oil reservoir at the base of the pump

• External cooling jacket or fan assembly attached to the pump body

• Jet assembly: a hollow assembly with nozzle(s)

• Cold Cap assembly attached to the inlet

• Inlet and exhaust outlet

• Electrical heater assembly

• Low vapor pressure oil (Use only the Beckman Coulter recommended silicone diffusion pump oil.)

Schematic of a diffusion pump

During normal operation of a Beckman Coulter ultracentrifuge, the vacuum pump runs first. The diffusion pump fan assembly operates. The oil reservoir heater is activated depending on the vacuum level, heating the oil until it boils. The hot oil vapor is then released through the jet nozzles, expanding as it moves from high to low pressure. As gas molecules (air and water) enter the diffusion pump inlet, they collide with the heavy oil vapor stream and are directed towards the pump's exhaust outlet by the vacuum pump. The supersonic vapor jet of diffusion pump oil then strikes the cooled diffusion pump wall, causing the oil vapor to condense and flow back into the oil reservoir to begin the cycle anew.

To achieve the necessary vacuum conditions in ultracentrifuges, the diffusion pump must be used in series with a vacuum pump. The diffusion pump cannot be exposed to the atmosphere while the oil is hot due to potential oil degradation and vacuum system contamination.

The lowest pressure achievable in the ultracentrifuge vacuum systems is partly limited by the presence of oil vapor from the diffusion pump. This occurs due to "backstreaming" of oil vapor molecules from the diffusion pump nozzle towards the inlet, as well as "back migration" caused by re-evaporation of oil at the top of the pump body.

The amount of backstreaming can be reduced by optimizing the pump nozzle design and using cold caps, traps, or baffles above the pump body. Controlling the temperature of the pump housing and the baffle system can also reduce back migration.

While this backstreaming and back migration is not a significant issue for typical preparative ultracentrifuges, the oil vapor can be problematic for the optical systems of analytical ultracentrifuges.

The diffusion pump is a low-maintenance component, with no mechanical moving parts. However, it is important to perform regular maintenance to ensure optimal performance. Here are the recommended maintenance tasks:

Check the diffusion pump oil regularly and change it if it appears light to dark brown in color.

  1. Remove the jet assembly once a year for inspection and cleaning.

  2. Inspect the oil heater and check its resistance.

  3. Inspect and clean the cooling fan and verify its operation.

Here are some troubleshooting tips for the diffusion pump: 

Oil Heater Assembly - If the heater is burned out, the system will only achieve a normal mechanical pump vacuum level (less than 100 microns). The reason for heater failure may be due to loss of oil, hours of operation, or loss of cooling.

Loss of Fluid - Normally, loss of fluid is due to a leak in the vacuum system.

Cooling Fan Malfunction - This causes a loss of diffusion pump cylinder cooling.

Loss of Vacuum - The system will initially achieve the correct vacuum. As the pump cylinder heats up, the vapor no longer condenses, and loss of vacuum occurs.

Diffusion Pump Fluid Contamination - If the drive oil or liquid from the rotor spills into the diffusion pump, it may cause the pump to lose effectiveness.

PMC Check - Check for the correct oil level, heater function, and air flow.

Normal Loss of Vacuum - With a chamber vacuum of 150 microns or better, it is normal to see a slight loss of vacuum immediately after turning on the diffusion pump heater.

Vacuum System Troubleshooting: Finding and Fixing Leaks:

  1. Check for obvious issues such as cracked hoses, tubing, and foreign material on "o" rings.

  2. Confirm the mechanical pump is working properly by measuring the vacuum level at the pump inlet, which should be less than 20 microns. Replace vacuum pump oil on a regular basis.

  3. Be aware that removing moisture from a vacuum system can take time, and during this period the vacuum level will remain at approximately 500 microns.

  4. For larger leaks, turn off power and listen for a hissing sound to detect the source of the leak.

  5. Try to isolate sections of the vacuum system to narrow down potential leak locations.

  6. Use acetone to identify leaks: if you're able to obtain a vacuum of 300 microns or less, squirt acetone on possible leak areas. The vacuum indicator will exhibit a loss of vacuum at the area where the leak is present.

  7. Ensure the diffusion pump is functional by following proper maintenance and troubleshooting procedures.