Contact Us

Got Questions? We're Here to Help! Contact Us.


California
United States

+1 (510) 899-1340

Welcome to Service AI, your trusted source for accessories, spare parts, and certified pre-owned laboratory instrumentation tailored for scientists and startups. We provide reliable, current equipment at competitive prices, along with essential support for instruments no longer made or backed by their original manufacturers.

Our expertise spans leading brands like Beckman Coulter and Agilent, and we are committed to high-quality standards through certified maintenance, repair, and professional installation services. We believe our installation services provide the confidence and assurance you need when purchasing instrumentation from a reliable and trustworthy source.

At Service AI, we don't conduct the science—we empower scientists with reliable laboratory instrumentation and spare parts that are guaranteed to perform. We trust in scientists' expertise to determine the instrumentation they need to advance their research.

With more than 25 years of experience, our dedicated team prides itself on delivering only the highest quality products. Every instrument we offer has been thoroughly tested and certified by manufacturer-trained engineers. We ensure the functionality of our products with a Full Performance Guarantee and provide proof of performance documentation when applicable.

At Service AI, we also recognize the importance of meticulous shipping and safe delivery. Our products are handled with the utmost care to ensure they arrive in perfect condition. For those who require it, we offer professional installation services to help integrate our equipment seamlessly into your operations.

Purchase our reliable scientific laboratory equipment, and we’ll provide seamless installation service worldwide for your startup or research needs.

Advance your scientific endeavors with confidence: rely on us for recertified, trustworthy instruments at reasonable prices. It’s a smart choice made simple!

Is your Beckman Coulter LS 6500 failing efficiencies?

Beckman LS Guide

Is your Beckman Coulter LS 6500 failing efficiencies?

Service AI

Acceptable values meet these system specifications:  

  • 3H: ≥ 59% in wide open window

  • 14C: ≥ 95%

The specifications mentioned above are relevant for Beckman calibrated standards that are traceable to the National Bureau of Standards. To ensure accurate results, it is imperative to calibrate the LS counter properly.

To ensure accurate and consistent results, it is essential to intimately mix any sample with the cocktail according to Beckman. Low efficiency and unpredictable results can occur if this step is not followed. Additionally, dirty detector or reflector housing, as well as aged PMTs, can also cause low efficiency reports. One can easily identify dirty reflector housing by visually inspecting it; a clean and reflective internal paint maximizes counter efficiencies, whereas a dark and dirty coating indicates contamination. Replacing the reflector housing is a simple and cost-effective solution to resolve efficiency issues in Beckman LS6500 counters.

To detect possible radioactive contamination and check overall performance, it is recommended to periodically conduct a background count and efficiency check following the procedure below.

Set up a rack with the supplied 3H, 14C, and unquenched background standards, followed by three or four of your own samples that are giving questionable results.

Set up a User Program as follows:

Count Time: 5 minutes

Isotope 1: 3H

Isotope 2: Wide

Data Calculation: CPM

Count Sample: 3 times

H# (if installed): ON

Lum-Ex: ON

3H and 14C precision: 1 %

3H Counting Efficiency:

For 3H it will be necessary to correct for the radioactive decay since the samples were prepared. This can be done by using the half-life table. The 3H standard will have the date of calibration and the dpm on that date printed on the vial. Calculate t (current date minus date of calibration. in years].

Divide t by the half-life T (12.35 years for 3H). Find this value in the t/T column of the half-life table. Use the corresponding fraction remaining value to multiply the standard DPM. The result is the DPM remaining on the selected date.

Use the CPM for Isotope I for the 3H standard, and the CPM for Isotope I for the background sample. The background for this calculation is insignificant compared to the CPM of the standard but is given for completeness. Calculate the 3H efficiency from the equation:

3-23-2021 8-53-40 PM.png

14C Counting Efficiency:

For the 14C efficiency. use the 14C CPM for Isotope 2, Wide. Calculate the efficiency as described for 3H.

Background:

As per Beckman, the background count in the Wide window should be less than 60 CPM. The background in an LS counter is influenced by the noise within the system (phototubes. electronics). Naturally occurring radioactivity, and cosmic rays. The background also varies with the size and type of vial (plastic has lower background than glass), and the altitude (about a 7% increase in background for each 1000 feet above sea level).

If any results fail to meet these values, check the following:  

  1. Inspect the sample vials to ensure they are securely sealed.

  2. Ensure the instrument is calibrated and set up properly.

  3. Verify that you have accurately corrected for decay using the Half-life Tables, and that your computations are correct.

  4. Replace LS6500 Elevator Cap.

  5. Inspect the detector reflector housing. If it looks dirty, replace the reflector housing.